Manufactur standard Cas No.:3327-22-8 - Medical folding nanofiber protective mask – E.Fine
Manufactur standard Cas No.:3327-22-8 - Medical folding nanofiber protective mask – E.Fine Detail:
folding nanofiber protective mask
Main raw material ,Structure and composition:
Mask (sterility) It is composed of mask body, nose clip and mask belt. Mask body is divided : surface layer, filtration layer , fine nano filtration layer , bottom layer,Skin friendly layer , total five layers.
The surface , bottom layer , and skin friendly layer of mask is Polypropylene spunbonded nonwoven fabric , filtration layer is Polypropylene melt-sprayed nonwoven fabric, fine nano filtration layer is nanofiber membrance , the nose clip is made of a plastic material,The mask belt is made of polyester and spandex yarn.
Application:
This standard is applicable to the medical working environment, filter the particulate matter in the air, block droplets, blood, body fluids, secretions and other self-absorption filter medical respirator.
Main technical indicators of products:
Filtration efficiency
When the gas flow rate is 85L/min, the filtration efficiency of the mask for non-oily particles should meet the requirements in table
sheet 1 ,Filtration efficiency class % |
|
One |
≥95 |
two |
≥99 |
Three |
≥99.97 |
Airflow resistance
For gas flow of 85L/min, the inspiratory resistance of the mask should not exceed 343.2pa (35mmH2O)
synthetic blood penetration
Spray 2mL of synthetic blood into the mask at a pressure of 10.7kpa (80mmHg). No infiltration should occur inside the mask
Surface moisture resistance
The water level on the outer surface of the mask should not be lower than the level 3 in GB/ t4745-1997
Microbiological indicator
The mask should meet the requirements of the microbial indicators in gb15979-2002, as shown in table 2
Sheet 2 Microbiological index of mask
Total number of bacterial colonies CFU/G |
coli group |
pseudomonas aeruginosa |
staphylococcus aureus |
hemolytic streptococcus |
Total number of fungal colonies |
≤200 |
undetectable |
undetectable |
undetectable |
undetectable |
≤100 |
Residual ethylene oxide
For masks sterilized by ethylene oxide, the residue of ethylene oxide should not exceed 10 g/g
flame resistance
The materials used should not be inflammable. The afterburn time should not exceed 5s
Skin irritation
The score for primary irritation of mask material should not exceed
Sterilization method: ethylene oxide sterilization. Residual ethylene oxide (g/g) ≤10
Standard:GB19083-2010
Product detail pictures:
Related Product Guide:
17 applicants chosen to move forward in Aquaculture Innovation Challenge | Livestock Betaine
Worldwide Probiotics in Animal Feed Market Demand Analysis 2019 | Organic Poultry Feed
We goal to see good quality disfigurement within the manufacturing and provide the most effective support to domestic and overseas shoppers wholeheartedly for Manufactur standard Cas No.:3327-22-8 - Medical folding nanofiber protective mask – E.Fine , The product will supply to all over the world, such as: Chile, Curacao, Niger, Due to our strict pursues in quality, and after-sale service, our product gets more and more popular around the world. Many clients came to visit our factory and place orders. And there are also many foreign friends who came for sight seeing, or entrust us to buy other stuff for them. You are most welcome to come to China, to our city and to our factory!
By Dana from Georgia - 2017.08.21 14:13
Wide range, good quality, reasonable prices and good service, advanced equipment, excellent talents and continuously strengthened technology forces,a nice business partner.
By Tyler Larson from Philippines - 2017.09.22 11:32